Calibrating pH and EC probes
pH and EC probes drift. Calibration is what turns a $10 probe into a trustworthy instrument — and Verdalio's ML layer only extends good calibration, it does not replace it.
What you need
- pH 4.01 and 7.01 buffer solutions
- EC 1.413 mS/cm standard solution
- Distilled water for rinsing between solutions
Calibrate pH (two-point)
- Rinse the probe in distilled water.
- Place it in pH 7.01 buffer; let the reading settle; confirm the 7.01 point.
- Rinse; place in pH 4.01 buffer; confirm the 4.01 point.
- Rinse and return to your solution.
Calibrate EC
Rinse, place the probe in the 1.413 mS/cm standard, let it settle, confirm. Always measure EC in mS/cm, not PPM — PPM conversion varies by manufacturer and will silently mislead you.
Operator discipline
- Calibrate pH probes every 14–28 days, EC probes every 2–4 weeks.
- Store pH probes in KCl storage solution between uses — never dry, never in distilled water.
- Clean probes monthly; see Cleaning & Sanitation.
Then let the ML layer help
Once a probe is calibrated, the hub's on-device model tracks its ongoing drift and corrects readings between manual calibrations. Cheap sensors, smart software.